Digital Conductivity Meter 601, 602 & 611

Features

  • Microcontroller Based
  • Highly Stable & Accurate
  • Auto Ranging Facility
  • Cell Constant Measurement and Adjustment Facility
  • 3½ Digit Display
  • Automatic Temperature Compensation Facility (601 & 602 only)
model-601Nw

EI’s Digital Conductivity Meters are reliable and accurate test instruments for measurement of Conductivity of aqueous solutions. They measure Conductivity and TDS in five ranges. The Resolution is 0.1 µS/cm and 0.1 ppm respectively, in the lowest range. The cell constant is flashed on a digital display & can be adjusted from the front panel.

These instruments are ideal for monitoring salt contents in water and soluble salts in soils. They are extremely useful instruments for agriculture and soil analysis labs, swimming pools, fertilizer plants, petroleum refineries, textile plants etc.

Specifications

Model 601 602 611
Display 3½ Digit LED
Measurement Conductivity & Temperature Conductivity & Temperature Conductivity
Ranges Five Ranges
0 – 200 µS/cm
0 – 2 mS/cm
0 – 20 mS/cm
0 – 200 mS/cm
0 – 1000 mS/cm
Six Ranges
0 – 20 µS/cm
0 – 200 µS/cm
0 – 2 mS/cm
0 – 20 mS/cm
0 – 200 mS/cm
0 – 1000 mS/cm
Five Ranges
0 – 200 µS/cm
0 – 2 mS/cm
0 – 20 mS/cm
0 – 200 mS/cm
0 – 1000 mS/cm
Range Selection Automatic in 5 Ranges Automatic in 6 Ranges Automatic in 5 Ranges
Temperature Range 0 to 100 °C 0 to 100 °C 0 to 100 °C
Accuracy
Cond.
Temp.
±0.5 % FS ± 1Digit
Temp: ±0.1°C ± 1Digit
±0.5 % FS ± 1Digit
±0.1°C ± 1Digit
±0.5 % FS ± 1Digit
Temperature
Compensation
Auto : 0 to 50 °C
Manual : 0 to 50 °C
Auto : 0 to 50 °C
Manual : 0 to 50 °C
Manual: 0 to 50 °C
Cell constant Adjustable on Digital Display
Measuring Cell Platinum DIP Type
Auto Cell Constant Measurement Available Available Available
Resolution
Cond.
Temp.
0.1 µS/cm
0. 1°C
0.01 µS/cm
0. 1°C
0.1 µS/cm
Power 12V DC using 230V + 10% AC Adapter
Dimensions 270x202x80 mm (lxbxh)
Weight 0.8 Kg approx.
Accessories Conductivity Cell, Operation Manual, Dust Cover, Cell Stand, Temperature Probe (with models 602 and 601)

Note : All Conductivity Meters are supplied with TDS Cell having cell constant of 1.0 approximately. Specifications are subject to change without notice due to continuous developments.